The Bionode: A Closed-Loop Neuromodulation Implant

Published:18 February 2019Publication History
Skip Abstract Section

Abstract

Implantable closed-loop neuromodulation devices for use in long-term chronic studies in a lab or clinical trial are expensive to acquire and difficult to modify for specific use cases. This article documents the design and fabrication of a wireless implantable device using only commercially available off-the-shelf (COTS) components. This device, called the Bionode, can record and transmit up to four channels of biopotential data while simultaneously providing biphasic constant-current stimulation. The Bionode is a viable, low-cost, reusable, and easily modifiable research tool with clinical implications that has gained widespread use in various research projects at Purdue University.

References

  1. D.-B. S. f. P. S. D. S. Group. 2001. Deep-brain stimulation of the subthalamic nucleus or the pars interna of the globus pallidus in Parkinson's disease. N. Engl. J. Med. 2001, 345 (2001), 956--963.Google ScholarGoogle Scholar
  2. D. M. Labiner and G. L. Ahern. 2007. Vagus nerve stimulation therapy in depression and epilepsy: Therapeutic parameter settings. Acta. Neurol. Scand. 115, 1 (2007), 23--33.Google ScholarGoogle ScholarCross RefCross Ref
  3. H. A. Sackeim. 2001. Vagus nerve stimulation (VNSTM) for treatment-resistant depression: Efficacy, side effects, and predictors of outcome. Neuropsychopharmacology 25, 5 (2001), 713--729.Google ScholarGoogle ScholarCross RefCross Ref
  4. J. Luigjes. 2012. Deep brain stimulation in addiction: A review of potential brain targets. Mol. Psychiatry 17, 6 (2012), 572.Google ScholarGoogle ScholarCross RefCross Ref
  5. F. Sun, M. Morrell, and R. Wharen. 2008. Responsive cortical stimulation for the treatment of epilepsy. Neurotherapeutics 5, 1 (2008), 68--74.Google ScholarGoogle ScholarCross RefCross Ref
  6. M. J. Kane, P. P. Breen, F. Quondamatteo, and O. L. Gearoid. 2011. BION microstimulators: A case study in the engineering of an electronic implantable medical device. Med. Eng. Phys. 33, 1 (2011), 7--16.Google ScholarGoogle ScholarCross RefCross Ref
  7. M. Weiergräber, M. Henry, J. Hescheler, N. Smyth, and T. Schneider. 2005. Electrocorticographic and deep intracerebral EEG recording in mice using a telemetry system. Brain Res. Protocols 14, 3 (2005), 154--164.Google ScholarGoogle ScholarCross RefCross Ref
  8. H. Kassiri et al. 2016. Battery-less tri-band-radio neuro-monitor and responsive neurostimulator for diagnostics and treatment of neurological disorders. IEEE J. Solid-State Circ. 51, 5 (2016), 1274--1289.Google ScholarGoogle ScholarCross RefCross Ref
  9. Q. Xu, D. Hu, B. Duan, and J. He. 2015. A fully implantable stimulator with wireless power and data transmission for experimental investigation of epidural spinal cord stimulation. IEEE Trans. Neural. Syst. Rehabil. Eng. 23, 4 (2015), 683--92.Google ScholarGoogle ScholarCross RefCross Ref
  10. R. G. Dennis, D. E. Dow, and J. A. Faulkner. 2003. An implantable device for stimulation of denervated muscles in rats. Medical Eng. Phys. 25, 3 (2003), 239--253.Google ScholarGoogle ScholarCross RefCross Ref
  11. S. G. Ewing, W. J. Lipski, A. A. Grace, and C. Winter. 2013. An inexpensive, charge-balanced rodent deep brain stimulation device: A step-by-step guide to its procurement and construction. J. Neurosci. Methods 219, 2 (2013), 324--330.Google ScholarGoogle ScholarCross RefCross Ref
  12. H. Kassiri et al. 2017. Closed-loop neurostimulators: A survey and a seizure-predicting design example for intractable epilepsy treatment. IEEE Trans. Biomed. Circ. Syst. 99 (2017), 1--15.Google ScholarGoogle Scholar
  13. R. Schaumann. 2010. Design of Analog Filters (2nd ed.). Oxford University Press, NY. Google ScholarGoogle ScholarDigital LibraryDigital Library
  14. R. F. Coughlin. 1991. Operational Amplifiers and Linear Integrated Circuits (4th ed.). Prentice Hall, Englewood Cliffs, NJ.Google ScholarGoogle Scholar
  15. T. August, M. Harvey, P. Lightfoot, D. Kilbey, T. Papadopoulos, and P. Jepson. 2015. Emerging technologies for biological recording. Biol. J. Linnean Soc. 115, 3 (2015), 731--749.Google ScholarGoogle ScholarCross RefCross Ref
  16. S. Nag and N. V. Thakor. 2016. Implantable neurotechnologies: Electrical stimulation and applications. Med. Biol. Eng. Comput. 54, 1 (2016), 63--76.Google ScholarGoogle ScholarCross RefCross Ref
  17. A. C. Thompson, P. R. Stoddart, and E. D. Jansen. 2014. Optical stimulation of neurons. Curr. Mol. Imag. 3, 2 (2014), 162--177.Google ScholarGoogle ScholarCross RefCross Ref
  18. S. T. Lee, P. A. Williams, C. E. Braine, D. T. Lin, S. W. M. John, and P. P. Irazoqui. 2015. A miniature, fiber-coupled, wireless, deep-brain optogenetic stimulator. IEEE Trans. Neural Syst. Rehab. Eng. 23, 4 (2015), 655--664.Google ScholarGoogle ScholarCross RefCross Ref
  19. J. E. Arle. 2011. Essential Neuromodulation. Elsevier Science, Burlington.Google ScholarGoogle Scholar
  20. D. R. Merrill, M. Bikson, and J. G. R. Jefferys. 2005. Electrical stimulation of excitable tissue: Design of efficacious and safe protocols. J. Neurosci. Methods 141, 2 (2005), 171--198.Google ScholarGoogle ScholarCross RefCross Ref
  21. AN-1515 A Comprehensive Study of the Howland Current Pump. 2013. Technical report, Texas Instruments.Google ScholarGoogle Scholar
  22. H. M. Brace, J. G. R. Jefferys, and J. Mellanby. 1985. Long-term changes in hippocampal physiology and leaning ability of rats after intrahippocampal tetanus toxin. J. Physiol. 368 (1985), 343--57.Google ScholarGoogle ScholarCross RefCross Ref
  23. D. Andreuccetti, R. Fossi, and C. Petrucci. 1997. An internet resource for the calculation of the dielectric properties of body tissues in the frequency range 10 Hz--100 GHz. Retrieved from http://niremf.ifac.cnr.it/tissprop/.Google ScholarGoogle Scholar
  24. W. L. Stutzman. 1981. Antenna Theory and Design. Wiley, NY.Google ScholarGoogle Scholar
  25. C. A. Balanis. 2005. Antenna Theory: Analysis and Design (3rd ed.). Wiley-Interscience, Hoboken, NJ. Google ScholarGoogle ScholarDigital LibraryDigital Library
  26. D. M. Pozar. 2012. Microwave Engineering (4th ed.). Wiley, Hoboken, NJ.Google ScholarGoogle Scholar
  27. A. Ben Amar, A. B. Kouki, and H. Cao. 2015. Power approaches for implantable medical devices. Sensors (Basel) 15, 11 (2015), 28889--914.Google ScholarGoogle ScholarCross RefCross Ref
  28. H. Mei and P. P. Irazoqui. 2014. Miniaturizing wireless implants. Nature Biotechnol. 32, 10 (2014), 1008--1010.Google ScholarGoogle ScholarCross RefCross Ref
  29. A. Kim and Purdue University. Electrical and Computer Engineering: Acoustically Powered Wireless Medical Implants; part 1, Electronic Resource. Purdue University.Google ScholarGoogle Scholar
  30. H. Mei, K. Thackston, R. Bercich, J. Jefferys, and P. Irazoqui. 2016. Cavity resonator wireless power transfer system for freely moving animal experiments. IEEE Trans. Biomed. Eng. 99 (2016), 1--1.Google ScholarGoogle Scholar
  31. H. Mei, Y. W. Huang, K. A. Thackston, and P. P. Irazoqui. 2016. Optimal wireless power transfer to systems in an enclosed resonant cavity. IEEE Antennas Wireless Propagat. Lett. 15 (2016), 1036--1039.Google ScholarGoogle ScholarCross RefCross Ref
  32. K. A. Thackston and Purdue University. Biomedical Engineering: Optimization of Wireless Power Networks for Biomedical Applications; part 1, Electronic Resource. Purdue University.Google ScholarGoogle Scholar
  33. D. Ha, T.-C. Lee, D. J. Webery, and W. J. Chappell. 2014. Power distribution to multiple implanted sensor devices using a multiport bandpass filter (BPF) approach. In IEEE MTT-S International Microwave Symposium (IMS'14). IEEE, 1--4.Google ScholarGoogle Scholar
  34. T. C. Beh, M. Kato, T. Imura, S. Oh, and Y. Hori. 2013. Automated impedance matching system for robust wireless power transfer via magnetic resonance coupling. IEEE Trans. Industr. Electron. 60, 9 (2013), 3689--3698.Google ScholarGoogle ScholarCross RefCross Ref
  35. Jesse P. Somann. 2018. A Study of Techniques of Vagus Nerve Stimulation for Treatment of Inflamation. PhD dissertation, Purdue University.Google ScholarGoogle Scholar
  36. G. Jiang and D. D. Zhou. 2009. Technology advances and challenges in hermetic packaging for implantable medical devices. In Implantable Neural Prostheses 2. Springer, 27--61.Google ScholarGoogle Scholar
  37. F. Boeser, J. S. Ordonez, M. Schuettler, T. Stieglitz, D. T. T. Plachta, and IEEE. 2015. Non-hermetic encapsulation for implantable electronic devices based on epoxy. In Proceedings of the 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 809--812.Google ScholarGoogle ScholarCross RefCross Ref
  38. C. K. Bjune et al. 2015. Package architecture and component design for an implanted neural stimulator with closed-loop control. In Proceedings of the 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 7825--7830.Google ScholarGoogle ScholarCross RefCross Ref
  39. P. Wang et al. 2015. Long-term evaluation of a non-hermetic micropackage technology for mems-based, implantable pressure sensors. In Proceedings of the 18th International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers). 484--487.Google ScholarGoogle ScholarCross RefCross Ref
  40. L. Lonys et al. 2015. Silicone rubber encapsulation for an endoscopically implantable gastrostimulator. Med. Biol. Eng. Comput. 53, 4 (2015), 319--329.Google ScholarGoogle ScholarCross RefCross Ref
  41. K. Stokes. 2010. The biocompatibility and biostability of new cardiovascular materials and devices. In Implantable Neural Prostheses 2: Techniques and Engineering Approaches, D. Zhou and E. Greenbaum (Eds.). Springer, NY, 1--26.Google ScholarGoogle Scholar
  42. Q. Xu, J. Li, W. Han, and H. Zhou. 2011. A fully implantable stimulator with wireless power and data transmission for experimental use in epidural spinal cord stimulation. In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE, 7230--7233.Google ScholarGoogle Scholar
  43. A. Lay-Ekuakille, G. Griffo, A. Massaro, F. Spano, and G. Gigli. 2016. Experimental characterization of an implantable neuro-packaging for EEG signal recording and measurement. Measurement 79 (2016), 321--330.Google ScholarGoogle ScholarCross RefCross Ref
  44. A. Kalaijakis. 2011. Medical device interconnects must address harsh environments and durability challenges. Electron. Compon. News 55, 7 (2011), 20--20.Google ScholarGoogle Scholar
  45. S. Grob, P. A. Tass, and C. Hauptmann. 2016. Capacitive feedthroughs for medical implants. Front. Neurosci. 10 (2016).Google ScholarGoogle Scholar
  46. R. A. Stevenson. 1998. Feedthrough EMI filter with ground isolation for cardiac pacemakers and implantable cardioverter defibrillators. In Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Parts 1--6: Biomedical Engineering Towards the Year 2000 and Beyond, vol. 20. H. K. Chang and Y. T. Zhang (Eds.). 3319--3323.Google ScholarGoogle ScholarCross RefCross Ref
  47. M. J. Kane, P. P. Breen, F. Quondamatteo, and G. ÓLaighin. 2011. BION microstimulators: A case study in the engineering of an electronic implantable medical device. Med. Eng. Phys. 33, 1 (2011), 7--16.Google ScholarGoogle ScholarCross RefCross Ref
  48. J. R. Lachapelle et al. 2016. An implantable, designed-for-human-use peripheral nerve stimulation and recording system for advanced prosthetics. In Proceedings of the IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC’16). IEEE, 1794--1797.Google ScholarGoogle ScholarCross RefCross Ref

Index Terms

  1. The Bionode: A Closed-Loop Neuromodulation Implant

        Recommendations

        Comments

        Login options

        Check if you have access through your login credentials or your institution to get full access on this article.

        Sign in

        Full Access

        PDF Format

        View or Download as a PDF file.

        PDF

        eReader

        View online with eReader.

        eReader

        HTML Format

        View this article in HTML Format .

        View HTML Format
        About Cookies On This Site

        We use cookies to ensure that we give you the best experience on our website.

        Learn more

        Got it!